INSTALLATION & OPERATION MANUAL MODEL SP1200 Flo-Calibration System DOC#: MN-1200
Enter the trailer totalizer reading noted at the start of the prover. Press “ENTER” key. Enter the trailer totalizer reading noted at the end
Reprove the trailer. Compare the prover totalizer reading to the trailer totalizer. Truck % error should be within specified tolerance. Re
©2009 Pub. No. MN-1200-Rev (09/09)
METER PROVER OPERATION-COMPREHENSIVE WARNING Never drop any component of the prover system-carefully place each component in desired location.
INITIAL SET-UP MODE Hold down “RESET” key and turn power “ON” The prover “SET-UP” menu will be displayed. Press “PRINT” key Set-up ticket will
Select “1-Clock” by pressing “1” key. Enter the correct hour (1-12). The number entered will be displayed on the second line of the display. Pr
“SET-UP” menu will be displayed: “2-RTD” should not be selected. This is a factory calibration function. Select “3-Factor” by pressing “3” key
Select “4=More” by pressing “4” key. “MORE” menu will be displayed. Select “1-Meter” by pressing “1” key. Press “CLEAR/FEED” key
Select “3 - Tractr” by pressing “3” key. Press “CLEAR/FEED” key. Enter Tractor number, Press “ENTER” key. “MORE” menu will again be displayed
SET-UP OPERATING MODE SP1200 will alternate these 2 displays until “RESET” is pressed. Correct product must be displayed in the upper left hand
Cool down is verified by the temperature display. The temperature displayed along with a small “c” indicates the prover piping is cooled down. The t