INSTALLATION & OPERATION MANUAL IT275N REV. A Remote Totalizer & Rate Indicator DOC#: MN-275N-A.doc
Sponsler, Inc. IT275N Remote Totalizer & Rate Indicator pg. 10 DOC#: MN-275N-A Example #8: Actual = 50 Displayed = 52 C.F. = .4000 New C.
Sponsler, Inc. IT275N Remote Totalizer & Rate Indicator pg. 11 DOC#: MN-275N-A Typical Output Configuration – Although the IT27
Sponsler, Inc. IT275N Remote Totalizer & Rate Indicator pg. 12 DOC#: MN-275N-A MISCELLANEOUS Battery Replacement The battery is located
Sponsler, Inc. IT275N Remote Totalizer & Rate Indicator pg. 13 DOC#: MN-275N-A Dimensional Information FIGURE 6
Sponsler, Inc. IT275N Remote Totalizer & Rate Indicator pg. 14 DOC#: MN-275N-A
Sponsler, Inc. IT275N Remote Totalizer & Rate Indicator pg. 15 DOC#: MN-275N-A
©2009 Pub. No. MN-275N-A (09/09)
Sponsler, Inc. IT275N Remote Totalizer & Rate Indicator pg. 2 DOC#: MN-275N-A TABLE OF CONTENTS PAGE # SPECIFICATIONS ...
Sponsler, Inc. IT275N Remote Totalizer & Rate Indicator pg. 3 DOC#: MN-275N-A SPECIFICATIONS TEMPERATURE: Operating -40 to 85°C Storage –65 t
Sponsler, Inc. IT275N Remote Totalizer & Rate Indicator pg. 4 DOC#: MN-275N-A INTRODUCTION General The Model IT275N Industrial Remote Totaliz
Sponsler, Inc. IT275N Remote Totalizer & Rate Indicator pg. 5 DOC#: MN-275N-A INSTALLATION Inspection All units are completely assemble
Sponsler, Inc. IT275N Remote Totalizer & Rate Indicator pg. 6 DOC#: MN-275N-A CALIBRATION Sensitivity The sensitivity adjust R1 is locate
Sponsler, Inc. IT275N Remote Totalizer & Rate Indicator pg. 7 DOC#: MN-275N-A On the IT275F PCB Set S5@4, S6, 7, 8@0 (.4000) Set S4 i
Sponsler, Inc. IT275N Remote Totalizer & Rate Indicator pg. 8 DOC#: MN-275N-A Example 3: Assume a turbine has a K-Factor of 79.58 pulses per SC
Sponsler, Inc. IT275N Remote Totalizer & Rate Indicator pg. 9 DOC#: MN-275N-A C.F. = 84.7/250 = .3388 On the IT275I PCB: Set S3 #1 ‘ON’ (↓